Fluctuation measurements in a supersonic boundary layer implying in-situ calibration of the hot-wire anemometer
a:2:{s:4:"TEXT";s:36:"B. Lenz, U. Gaisbauer, and E. Krдmer";s:4:"TYPE";s:4:"text";}
Institut f
Pages: 625–630
Abstract
A calibration technique for the constant-temperature hot-wire anemometer is presented, which is based on traversing the probe through the boundary layer of a flat plate while simultaneously performing fluctuation measurements. The free stream Mach number was M = 2.54, and the Reynolds number Red, based on wire diameter, ranged from 9 to 23. A comparison of the sensitivity values obtained with the aid of such a calibration procedure
|